TOMOS Pipette Autoclaving instructions

TOMOS Pipette Autoclaving instructions
 
Please follow these instructions carefully in order to avoid damage to pipettes.
1.       All TOMOS EPP pipette and IPP pipette models can be autoclaved in one piece at 121 degree for 20 minutes. No special preparations are needed for autoclaving. You can use steam sterilization bags if needed.
2.       after autoclaving, the pipette must be cooled at room temperature for at least two hours before use.
3.   Check the calibration of the pipette after every autoclave treatment.