Maintenance of TOMOS Pipettes

Maintenance of Finnpipettes
 
Maintenance of TOMOS Pipettes
 
TOMOS Pipettes are easy to service in the laboratory using the tools provided with the pipette. All Finnpipettes come with detailed instructions on how to disassemble and assemble the pipette. Instructions for routine in-lab maintenance are also included. Instructions for use can also be found on our websites.

Short-term checking

At the beginning of each workday, the pipette should be checked for dust and dirt on the outside surfaces. Particular attention should be paid to the tip cone. No other solvents except 70% ethyl alcohol should be used to clean the pipette.

Long-term maintenance

If the pipette is used daily, it is recommended to check the calibration every five months.

General guidelines for cleaning the pipette:

1. Open the pipette.

2. Check for dust and dirt, and wipe clean.

3. Use only 70% ethyl alcohol to clean. The tip cone and tip ejector can be left in an alcohol bath overnight.

4. Grease the piston, O-ring and other cleaned parts with the silicon grease that comes with the pipette.

5. Assemble the pipette.

Please note: The calibration has to be checked after cleaning the pipette.

autoclavable pipettes resistance

TOMOS EasyPette Plus and InnoPette Plus models are full Autoclavable pipette.